Functional Description
47
4.6.3
Screen icons of the bus analysis mode
Selection of the configuration and display of the connections necessary for the analysis of the
selected bus.
SCOPIX IV proposes a set of bus configurations and connection diagrams. These files cannot be
deleted or modified, but can be copied and then modified. The .bus* file extension identifies
configurations that have been modified by the user. The user must select one of these files to be
able to start an analysis:
When a configuration file has been selected, the standard (or directive) and the probe connections
diagram are displayed.
Measurement tolerances
Display of the tolerances applied according to the standard or directive in force. You can change
these tolerances by clicking the value to be modified. The modifications are automatically recorded
in the .bus* file copied in the folder named "bus-limits". The measurements "tolerances" menu
contains: the min and max intervals of each measurement and the interval "of acceptability"
outside the interval of tolerances (as a percentage of the interval defined by the min and max
values).