BAM 1020-9800 Manual Rev W.docx
Page 85
BAM 1020 Settings Report
10/15/17 14:30:35
Station ID, 195
Serial Number, M8195
Firmware BAM, 3236-05 V3.12.0
Firmware 80350, 80353-03 R2.7.0
K, 01.000
BKGD, 0.0000
usw, 00.285
ABS, 00.844
Range, 1.000
Offset, -0.015
Clamp, -0.015
Conc Units, ug/m3
Conc Type, ACTUAL
Count Time, 4
Conc Error, FULL SCALE VALUE
Inlet Type, PM10
Cv, 01.000
Qo, 00.000
Flow Type, ACTUAL
Flow Setpt, 0016.7
Std Temp, 25
High Flow Alarm, 20
Low Flow Alarm, 10
Heat Mode, AUTO
Heat OFF (%), 20
RH Ctrl, YES
RH SetPt, 35
RH Log, YES
DT Ctrl, NO
DT SetPt, 99
DT Log, NO
BAM Sample, 50
MET Sample, 5
Cycle Mode, STANDARD
Fault Polarity, NORM
Reset Polarity, NORM
Maintenance, OFF
HJ 653, YES
EUMILRNFPDCT
111111111111
AP, 000150
Baud Rate, 9600
Printer Report, 2
e3, 00.000
e4, 15.000
Channel, 1, 2, 3, 4, 5, 6,
Sensor ID, 255, 255, 255, 255, 255, 18,
Channel ID, 255, 255, 255, 255, 255, 254,
Name, XXXXX, XXXXX, XXXXX, RH , XXXXX, AT ,
Units, XXX, XXX, XXX, % , XXX, C ,
Prec, 0, 0, 0, 0, 0, 1,
FS Volts, 01.000, 01.000, 01.000, 00.500, 01.000, 01.000,
Mult, 000001, 000001, 000001, 32.258, 000001, 0080.0,
Offset, 000000, 000000, 000000, -25.81, 000000, -030.0,
Vect/Scalar, S, S, S, S, S, S,
Inv Slope, N, N, N, N, N, N,
Calibration, Offset, Slope,
Flow, 0.000, 1.000,
AT, 0.000,
BP, 0.000,
RH, 0.000,
FT, 0.000,
QUERY, 18, CONC_A, Q_STD, Q_ACT, ERRORS, REF, FLOW, CV,
AT, BP, XXXXX, XXXXX, XXXXX, RH , XXXXX, AT , TIME, ERRORS,
CONC_S,
Daily Range, 01:00 - 24:00
Span Check, 1 HR
Ethernet Handshake, None
File 4 System Configuration (Settings) File Example