N:08 DATA QUERY FIELDS
01 TIME 02 CONC_A 03 FLOW
04 AT 05 BP 06 RH
07 REF 08 ERRORS 09
10 11 12
13 14 15
16 17
SAVE EXIT
The SETUP > QUERY Screen
The
N:
field sets how many values to include in the array. Up to 17 parameters can be
included. Increment the N value with the up/down arrow keys. For each increment of the N
value, another position in the array will be activated, starting with position 01.
Each position in the array can be changed to any desired parameter using the left/right keys
to select the position, and the up/down keys to scroll through the whole list of parameters. In
the example above, eight parameters are included and the shown parameters were selected
for each of the eight positions in the array. The available parameters are listed in the table
below:
Parameter
Description
CONC_A
Concentration value for the last sample period.
Q_STD
Sample flow volume in cubic meters at standard conditions.
Q_ACT
Sample flow volume in cubic meters at actual AT/BP conditions.
STAB
Stability measurement. Diagnostics use only. EU firmware only.
REF
Reference span membrane mass measurement in mg/cm2.
FLOW
Real time flow or average flow for the last sample.
CV
Flow coefficient of variability for the last sample period. (Standard
deviation divided by the mean)
AT
Average ambient temperature for the sample period.
BP
Average barometric pressure for the sample period.
ANALOG 1
Average of analog met sensor channel 1 (user defined channel).
ANALOG 2
Average of analog met sensor channel 2 (user defined channel).
ANALOG 3
Average of analog met sensor channel 3 (user defined channel).
ANALOG 4
Average of analog met sensor channel 4 (usually filter RH).
ANALOG 5
Average of analog met sensor channel 5 (usually Delta T).
ANALOG 6
Average of analog met sensor channel 6 (almost always AT).
CONC_S
PM10 concentration using standard conditions. Used in PM-coarse
systems where the coarse PM10 value is in actual conditions, but
the standard PM10 value is also needed for regular PM10 reporting
requirements. This parameter is only available in the PM10 master
unit of a coarse set
PM2.5
PM2.5 concentration from the slave unit in a PM-coarse. This
parameter is only available in the master unit of a coarse set.
PMc
PM-coarse (PM10 - PM2.5) concentration value. This parameter is
only available in the master unit of a coarse set.
TIME
Date and Time stamp for the sample period. Ignored for BH
protocol.
ERRORS
Decimal error codes of the 12 major error categories.