BAM 1020-9800 Manual Rev W.docx
Page 39
6
SETUP MENU DESCRIPTIONS
The BAM 1020 uses a comprehensive system of setup menus which contain all of the
settings and parameters needed to perform the measurement and operation of the BAM
1020. Most of these settings are set at factory default values. Some settings may be altered
by the operator. This section describes the SETUP menu in detail, and should be reviewed
when the instrument is first put into service. Once set, most of the values in the SETUP
menus will not need to be changed. The SETUP values will not be lost if the BAM 1020 is
unplugged or powered down.
WARNING: Some of the settings in the SETUP menus are unit-specific calibration
constants which must not be changed, or the accuracy and proper operation of the
BAM 1020 may be affected.
WARNING:
Entering the SETUP menu system requires stopping the sample cycle.
Older versions of firmware will not warn you before stopping the sample!
Press the SETUP soft-key to enter the menu as shown below. The Setup Menu provides a
choice of operations. Use the arrow keys to navigate to the desired field, then press the
SELECT soft-key to enter.
SETUP MODE SELECT
CLOCK SAMPLE CALIBRATE EXTRA
ERRORS PASSWORD INTERFACE SENSOR
HEATER QUERY REPORTS HJ 653
SELECT EXIT
The SETUP Menu
A brief description of each sub-menu in shown in the table below. Detailed information is
provided in the following subsections.
Menu
Settings
CLOCK
BAM 1020 real-time clock date and time settings.
SAMPLE
Range, Offset, Sample Time, Count Time, Conc Units, Avg Period, Unit ID, and RS-232 settings.
CALIBRATE
Factory Calibration Values, (C
v
, Q
0
, ABS,
sw, K, BKGD) Flow rate, Flow type, Conc Type.
EXTRA1
Low concentration clamp, e1
– e4, Rarely used.
ERRORS
Analog error selections, Flow limits, Pressure drop limit.
PASSWORD
Password change screen.
INTERFACE
Cycle Mode early/standard, alarm relay polarity.
SENSOR
Meteorological sensor scaling and configuration screens, Channels 1
– 6.
HEATER
RH and Delta-T set-points for Smart Heater. Only visible if Heater Control is set to AUTO.
QUERY
Configuration for the custom Query data output file, and the European Bayern-Hessen protocol.
REPORTS
Daily data report hours.
HJ 653
Data format compliance with Chinese HJ 653-2013 standard