The SENSOR Menu
CH:
Selects the channel to be viewed. The up/down arrow keys select the channel.
TYPE:
This is the parameter name. You can enter a name here by using the arrow
keys to scroll through the alphabet and other ASCII characters.
UNITS:
This is the measurement units label for the channel. You can enter a value here
by using the arrow keys to scroll through the alphabet and other characters.
PREC:
This is the precision field, which sets the number of available decimal places for
the Multiplier and Offset parameters.
MULT:
This is really the measurement range or the span of the sensor. If a baro sensor
has a range of 525 to 825 mmHg, then the MULT would be 300 (mmHg). If a
RH sensor has a range of 0 to 100%, then the MULT would simply be 100 (%).
OFFSET:
This is the range offset value, or the measurement value that the sensor
represents at 0.000V output. In the screen shown above, the AT sensor has a
0-1V output representing -50 to +50°C. So the MULT range is 100 (C) and the
offset is -50, because 0.000V from the sensor represents -50°C.
FS VOLT:
This is the full-scale voltage output of the sensor. The maximum voltage range
that can be supplied by the sensor. This value is usually going to be either
1.000 or 2.500 volts. 2.500 is the maximum setting for this field.
INV SLOPE:
This setting allows the channel to recognize a sensor with an inverse slope.
This is always set to
N
(no) except for use with thermistor temperature sensors
with resistance-only outputs.
VECT/SCALAR:
This value sets the averaging method.
S
(scalar) is used for all
measurements except wind direction, which uses
V
(vector).
MODE:
This field is toggled by pressing the ID MODE soft-key. The value can be set to
either MANUAL or AUTO ID. In MANUAL mode, the user can enter their own
setup parameters for the channel. AUTO ID mode is used with 500 series
sensors, and must be selected in order for the BAM 1020 to recognize the
sensor automatically.
NOTE:
Any manually set parameters for that channel will
be lost when changing to AUTO ID mode.
Channel 6 must be set to AUTO ID
for all BAM 1020 monitors equipped with an ambient temperature sensor.
SETUP CHAN PARAMS
CH TYPE UNITS PREC MULT OFFSET
06 AT C 1 0100.0 -050.0
SENSOR FS VOLT: 1.000
INV SLOPE:N VECT/SCALAR:S MODE:AUTO ID
SAVE ID MODE EXIT