BAM 1020-9800 Manual Rev W.docx
Page 26
3.10 The OPERATE Menus
Press OPERATE soft-key at the main menu to enter operate menu as shown below. This will
not interrupt the sample if already running.
The OPERATE Menu
The DOWN arrow can be used to set the Operation Mode from ON to OFF. This will simply
stop the measurement cycle, but will not power-down the BAM 1020.
NOTE:
Even if the
operator sets the Operation Mode to OFF, or the BAM 1020 stops itself due to an error,
it will still automatically set the mode back to ON at the top of the hour, and try to run a
new cycle!
The only ways to prevent the BAM 1020 from automatically starting an hourly
cycle are to power off the BAM 1020, leave the BAM 1020 in a TEST or SETUP menu, or
leave the pinch rollers latched in the UP position.
The OPERATE menu has three soft-key options for viewing the operating status and sensor
measurements while the BAM 1020 is operating: NORMAL, INST, and AVERAGE.
3.11 The NORMAL Operation Screen
Normal Mode is the primary operation screen which displays most of the important
parameters of the sample progress in one place, as shown below. Many operators leave their
BAM 1020 in the NORMAL screen whenever the BAM 1020 is operating, instead of the Main
menu.
The NORMAL Operation Menu
The LAST C value indicates the last concentration record, updated at the end of the cycle.
The LAST m value indicates the last measured value of the reference span membrane. The
value should be very close or equal to the expected span value (ABS). The other values are
instantaneous measurements.
11/15/2006 OPERATE MODE 14:13:07
= ON
= OFF
Operation Mode: ON
Status: ON
NORMAL INST AVERAGE EXIT
11/15/2009
Normal Mode
11:27:54
LAST C: 0.061 mg/m3
LAST m: 0.806 mg/cm2
STATUS: SAMPLING
Flow: 16.7 LPM
AMB P: 764 mmHg
TAPE P: 584 mmHg
RH: 27 %
Heater: OFF
Delta-T: 4.2 C
EXIT