1.7 BAM 1020 Specifications
PARAMETER
SPECIFICATION
Measurement Principle:
Particulate Concentration by Beta Attenuation.
U.S. EPA Designations:
PM
10
: EQPM-0798-122
PM
2.5
EQPM-0308-170
PM
2.5
EQPM-0715-266
PM
10-2.5
FEM: EQPM-0709-185
Standard Range:
0 - 1.000 mg/m
3
(0 - 1000
g/m
3
)
Optional Ranges:
0 - 0.100, 0.200, 0.250, 0.500, 2.000, 5.000, 10.000 mg/m
3
(special applications)
Accuracy:
Exceeds US-EPA Class III PM
2.5
FEM standards for additive and multiplicative bias.
Lower Detection Limit:
(2σ) (1 hour)
< 4.8
g/m
3
(< 4.0
g/m
3
typical) (8-minute count time)
Lower Detection Limit:
(2σ) (24 hour)
< 1.0
g/m
3
Measurement Cycle Time:
1 hour
Flow Rate:
16.70 liters/minute
Filter Tape:
Glass fiber filter
Span Check:
Nominally 800
μg/cm
2
Beta Source:
C-14 (carbon-14), 60 µCi ±15 µCi (< 2.22 X 106 Beq), Half-Life 5730 years.
Beta Detector Type:
Photomultiplier tube with scintillator.
Operating Temp. Range:
0° to +50°C
Ambient Temp. Range:
-40° to +55°C standard. Optional -50 °C temperature sensors available.
Ambient Humidity Range:
0 to 90% RH, non-condensing.
Humidity Control:
Actively controlled inlet heater module, 10% - 99% RH set point.
Approvals:
U.S. EPA, MCERTS, CE, NRC, TUV, CARB, ISO-9001.
Standard User Interface:
Menu-driven interface with 8x40 character LCD display and dynamic keypad.
Optional User Interface:
Graphic color touch screen display module, Model BX-970.
Analog Output:
Isolated 0-1 VDC output standard. 0-10V, 4-20mA, 0-16mA switch-selectable.
Serial Interface:
RS-232 2-way serial ports for PC or modem communications.
Printer Output:
Output-only serial port, data or diagnostic output to a PC or serial printer.
Telemetry Inputs:
Clock Reset (voltage or contact closure), Telemeter Fault (contact closure).
Alarm Contact Closures:
Data Error, Tape Fault, Flow Error, Power Failure, Maintenance.
Compatible Software:
Air Plus™, Comet™, MicroMet Plus
®
, HyperTerminal
®
, ProComm Plus
®
.
Error Reporting:
User-configurable. Available through serial port, display, and relay outputs.
Memory:
4369 records (182 days @ 1 record/hr). Extended memory Report Processor option
Power Supply:
100 - 230 VAC, 50/60 Hz. 0.4 kW, 3.4A max @110V. Not including shelter.
Weight:
24.5 kg (54 lbs) without external accessories.
Unit Dimensions:
H x W x D = 31cm x 43cm x 40cm (12.25
” x 17” x 16”).
Specifications may be subject to change without notice.