BAM 1020-9800 Manual Rev W.docx
Page 53
6.12
HJ 653 Setup Screen
– Chinese Data Formatting
The SETUP > HJ 653 screen is used to set data formatting to match the Chinese National
Standards on Environmental Protection document HJ 653-2013. When set to YES, the
concentration display resolution on all screens will show micrograms to the tenths position. It
will be shown in the format of “x.x ug/m3”.
The data report will be in one of two formats, depending on the concentration type selected
(see section 6.3). Flow must be set to ACTUAL in both circumstances.
If the concentration is also set to ACTUAL, it will be formatted in this manner:
Time,ConcA(ug/m3),QtotA(m3),XXXXX(XXX),XXXXX(XXX),XXXXX(XXX),RH(%),XXXXX(XXX),AT(C
),ConcS(ug/m3),QtotS(m3),BP(kPa),Flow(LPM),E,U,M,I,L,R,N,F,P,D,C,T
2017-10-15 16:40, 123.4, 0.700, 0, 0, 0, 39, 0, 24.3, 139.1,
0.621, 100.2, 16.69, 0,0,0,0,0,0,0,0,0,0,0,0
If the concentration is set to STANDARD, it will be formatted in this manner:
Time,ConcS(ug/m3),QtotA(m3),XXXXX(XXX),XXXXX(XXX),XXXXX(XXX),RH(%),XXXXX(XXX),AT(C
),ConcA(ug/m3),QtotS(m3),BP(kPa),Flow(LPM),E,U,M,I,L,R,N,F,P,D,C,T
2017-10-15 16:40, 139.1, 0.700, 0, 0, 0, 39, 0, 24.3, 123.4,
0.621, 100.2, 16.69, 0,0,0,0,0,0,0,0,0,0,0,0
Also, selecting YES will force the DYNAMIC RANGE setting (see section 6.11.2) to
EXTENDED.
If the HJ 653 option is set to OFF, the data report will be as shown in section 9.4.