The BAM 1020 Standard User Interface and Keypad
Soft Keys:
Directly beneath the display are four white buttons called “soft-keys” or “hot-keys”. These are
dynamic keys where the function changes in response to a menu option displayed directly
above each key on the bottom row of the display. Whatever menu option is displayed above
one of these keys is the function which that key will perform in that particular menu. These
are used throughout the entire menu system for a variety of functions. For example,
modifications made within a menu are usually not saved unless a SAVE soft-key is pressed.
EXIT is also another common soft-key function.
Arrow (Cursor) Keys:
The four red arrow keys are used to scroll up, down, left, and right to navigate in the menu
system, and to select items or change fields on the screen. The arrow keys are also often
used to change parameters or increment/decrement values in the menu system.
Contrast Key:
The key with a circular symbol on it is for adjusting the light/dark contrast on the LCD display.
Press and hold the key until the desired contrast is achieved. It is possible to over-adjust the
contrast and make the entire display completely blank or completely dark, so be careful to set
it to a visible level.
Function Keys F1 to F6:
The function keys serve as shortcuts to commonly used menu screens. The
F
keys are only
functional from the main menu screen, or for entering passwords. The factory default
password is F1, F2, F3, F4.
F1
“Current”:
This key is a shortcut to the OPERATE > INST screen, used to display
instantaneous data values being measured by the BAM 1020. See section 3.12.
F2
“Average”:
This key is a shortcut to the OPERATE > AVERAGE screen, used to
display the latest average of the data recorded by the BAM 1020. See Section 3.12.
1/30/2018 BAM 1020 PM2.5 FEM 16:08:29
LAST CONCENTRATION: 0.028 mg/m3
CURRENT FLOW: 16.7 LPM
STATUS: ON
FIRMWARE: 3236-05 3.13.0
SETUP OPERATE TEST
TAPE
F1
F2
F3
F4
F5
F6