BAM 1020-9800 Manual Rev W.docx
Page 68
7.10 PUMP Test Menu
– Manual Pump and Nozzle Tests
The TEST > PUMP screen is used to perform leak checks. It can also be used to manually
force the pump on and off, or to manually move the nozzle.
Note:
The BAM will regulate the
flow to the 16.70 L/min set point, but the flow rate shown on this screen is uncorrected and
always in standard conditions, even if the BAM operates in actual flow mode
. For this reason,
no flow audits or checks should be performed using this screen! Obsolete BAM 1020 units
with a manual flow valve were calibrated using this screen.
NOZZLE/PUMP EST MODE
NOZZLE:
FLOW: 16.67 SLPM PUMP: ON
MOVE NOZZLE PUMP ON PUMP OFF EXIT
The PUMP Test screen
The NOZZLE status value will indicate if the nozzle is currently UP (
) or DOWN (
). The
PUMP status indicates if the pump is turned ON or OFF. The FLOW value is the current flow
rate, displayed in standard liters per minute (25C) only.
The MOVE NOZZLE soft key can be used to force the nozzle up or down for testing
purposes. Elapsed time is about 5 seconds. If the pump is ON this operation is disabled.
The PUMP ON and PUMP OFF soft keys can be used to turn the vacuum pump on or off.
The nozzle will be lowered automatically if PUMP ON is pressed.
7.11 TAPE Test Menu
– Manual Filter Forward/Backward Tests
The TEST > TAPE menu allows the user to manually move the filter tape forwards or
backwards in increments of 12.5mm “windows”. This is useful for spooling up the first few
turns of a fresh roll of tape, to test the tape transport mechanism, or to change spots of tape
for flow or count tests. The nozzle will be automatically raised if necessary, and the tape will
take a couple of seconds to move each window.
The “X:” value is the number of windows moved in the last motion. This number will be
negative if the last move was backwards.
The FEED value is the number of tape windows you want to move. Use the arrow up/down
keys to select up to 10 windows at a time.
The FWD and BKWD soft key move the tape forward or backward by the current amount of
the FEED value.
7.12 DAC Test Menu
– Analog Output Test
The TEST > DAC screen is used to test the function of the analog output voltage and the
DAC (digital/analog converter) electronics. Use the up/down arrow keys to force the voltage