M
AVTMMTO210 Rev 7 Sept 2013
12
4.
Connect the Serial Cable (optional).
If the user chooses to use an
external printer or PC, using PowerDB Lite PC software, then connect
the serial cable between the MTO and the printer or PC at this time.
5.
Connect the Remote Control cable (optional).
If the user chooses to
operate the
Test Button
of the MTO from a remote distance then,
connect the RCC cable at this time.
6.
Connect Warning Beacon (optional).
If the user chooses to operate
the
Warning Beacon
then, connect the WBC cable at this time and place
the beacon in a conspicuous location.
7.
Connect Interlock Switch.
The MTO has an external interlock switch
feature. Connect the external Interlock Switch via banana jack
connection to the interlock input jacks on the front panel. The interlock
switch function is a fail open, therefore if not used, an interlock jumper
shall be placed in the jacks. If the circuit is open “IntLoc” will be
displayed at the middle LCD.
8.
Connect the I and V leads (to the MTO end only at this time).
With
the clamps disconnected from the UUT, connect the I and V cables to
the MTO at this time. Be sure that all plugs are fastened securely to the
MTO so they will not become loose even in the event of the operator
inadvertently tripping over the current leads.
MTO210 is programmed to read two voltage inputs at all times. If a
voltage input is not used, the high impedance input characteristics of the
input cause the unit to give readings that are not zero. If a jumper is
installed, the reading will be very low.
9.
Connecting to the Transformer.
When testing high-voltage
transformers, caution must be used at all times and all safety precautions
followed. Read, understand, and employ all safety precautions and circuit
connections described in Sections 2
Safety
and Section 3
Preparing for Test.
F
WARNING
Ensure that the transformer to be tested is completely
de-energized. Check every winding. Ensure that all
terminals of the transformer are disconnected from line
or load at the transformer. Connections to ground may
be left in place.
F
WARNING
For all testing as described herein, care shall be taken
to ensure any and all unused clamps shall be isolated
from each other, from ground, and from personnel.
Содержание MTO210
Страница 2: ......
Страница 3: ...Transformer Ohmmeter DC Winding Resistance Test Set MTO210...
Страница 8: ...AVTMMTO210 Rev 7 Sept 2013 iv M...
Страница 18: ...M AVTMMTO210 Rev 7 Sept 2013 10 M...
Страница 22: ...M AVTMMTO210 Rev 7 Sept 2013 14 M...
Страница 32: ...M AVTMMTO210 Rev 7 Sept 2013 24 M...
Страница 48: ...M AVTMMTO210 Rev 7 Sept 2013 40 M...
Страница 72: ...M AVTMMTO210 Rev 7 Sept 2013 64 M...
Страница 74: ...M AVTMMTO210 Rev 7 Sept 2013 66 M...