CODE Error Description
Possible Cause
241
Onboard 10v out of range, voltage too low
242
Onboard 28v out of range, voltage too low
243
Discharge temperature is out of range, too
high
Unit overheated, fans filters clogged, fans
not working, overuse, extremely hot test
conditions
244
VICOR temperature is out of range, too
high
Unit overheated, fans filters clogged, fans
not working, overuse, extremely hot test
conditions
245
VNOISE voltage out of range, too low.
246
VICOR temperature too much ripple
3.0
ADC_ERROR
301
ADC reading times out or ADC locks up
302
Internal error. Communication with ADC
failed.
303
ADC sample rate setting failed
304
ADC internal zero cal failed
305
ADC internal full-scale calibration failed
306
ADC cannot be restored back to normal
mode
307
ADC channel 1 selection failed
308
ADC channel 2 selection failed
309
ADC channel 3 selection failed
310
Too much noise/ripple on ADC readings
311
ADC failed, initialization
312
ADC failed, run time
313
ADC locks up, can not read