M
AVTMMTO210 Rev 7 Sept 2013
56
ERROR CODES
MTO comes with a wide range of error codes to assure unit is working properly
and track down problems in a systematic manner. When an error code occurs,
the first thing to do is acknowledge the error by pressing enter and repeating the
test. If the error persists, review the possible causes and determine if you can
resolve the problem on site. If the basic cause a ruled out, contact Megger for
additional assistance. Bold Error codes are quite common with the possible
causes defined.
CODE Error Description
Possible Cause
ESd
ESD is pushed down
Button is pushed down, internal cable is
disconnected, internal hardware failure
IntLoc
Interlock interface is open
Interlock Jumper removed, attached
interlock circuit broken, internal cable
disconnected
1XX
Ethernet and top panel errors (MTO210
ONLY)
101
No link between top panel and lower
acquisition board
102
Ethernet communication failed
110
Current selection switch error
111
Function Mode switch error
112
Storage Mode Switch errors
113
I2C BUS error
114
LCD I2C bus error
115
LED I2C bus error
116
MUX I2C bus error
117
EEPROM failure
2.0
Initialization error
201
ADC internal zero calibration failure
202
ADC internal full-scale calibration failure
Содержание MTO210
Страница 2: ......
Страница 3: ...Transformer Ohmmeter DC Winding Resistance Test Set MTO210...
Страница 8: ...AVTMMTO210 Rev 7 Sept 2013 iv M...
Страница 18: ...M AVTMMTO210 Rev 7 Sept 2013 10 M...
Страница 22: ...M AVTMMTO210 Rev 7 Sept 2013 14 M...
Страница 32: ...M AVTMMTO210 Rev 7 Sept 2013 24 M...
Страница 48: ...M AVTMMTO210 Rev 7 Sept 2013 40 M...
Страница 72: ...M AVTMMTO210 Rev 7 Sept 2013 64 M...