2 Installation and Operation
2-3
2-3-4 Input Voltage/Current
<750W to 1.2kW type>
・When inputting 100V/110V/115V/120V AC, input change short bar SHORT.
・When inputting 200V/220V/230V/240V AC, input change short bar OPEN.
・Do not touch input change short bar, because high Voltage is being generated on it.
MODEL
INPUT
VOLTAGE
phase
INPUT CURRENT
±10%
AC50/60Hz
Normal type
(Typ)
PFC type
(Typ)
OUTPUT
VOLTAGE
750W to 765W
115V
1φ
12A
――
230V
8A
1.1kW to 1.2kW
115V
1φ
19A
――
230V
11A
1.8kW to 2.1kW
220V
1φ
17A
――
3φ
10A
――
2.9kw to 3kW
220V
1φ
25A
――
3φ
14A
――
3.75kW to 4kW
220V
3φ
19A
15A
4.5kW to 5.1kW
220V
3φ
23A
16A
7.35kW 10V, 15V
220V
3φ
35A
25A
to
20V to 60V
220V
3φ
34A
25A
7.5kW Over 100V
220V
3φ
33A
25A
8kW
10V, 15V
220V
3φ
46A
36A
to
20V to 60V
220V
3φ
44A
34A
10.5kW Over 100V
220V
3φ
41A
32A
11.7kW to 12kw
220V
3φ
54A
40A
15kw
220V
3φ
68A
50A
-L(200V) 115V of above INPUT VOLTAGE / CURRENT table become 100V,and 230V become 200V.
Input current is about 115% of Input current.220V(3φ) become 200V. Input current is about
110% of Input current.
-L(220V) 115V of above INPUT VOLTAGE / CURRENT table become 110V,and 230V become 220V.
Input current is about 105% of Input current.
-L(240V) 115V of above INPUT VOLTAGE / CURRENT table become 120V,and 230V become 240V.
Input current is about 95% of Input current. 220V(3φ) become 240V. Input current is about
90% of Input current.
※ 1)Other input voltage than AC115V or AC230V are optional.
2)Input current is the values at rated input voltage
3)Models other than standard article should follow the contents of each specification.
4)Single phase input corresponds only to 750W to 3kW.
!
MUST
・
Verify that rated input voltage is not decreased more than 10% of rated input voltage
(ex. 180VAC for 200VAC) at maximum rated voltage.
・Verify that input voltage is not decreased more than 10% of rated input voltage when
operating with all the devices connected.
・Check the input voltage indication on the rear panel of the unit and input the correct
voltage. Inputting wrong voltage may damage the device.
CAUTION
!
MUST