6 Operating by Analog Remote Control
52
HARS series
6-6 Output Current Program by External Voltage
Confirm that it is set to analog remote control.
(See [6-4 Analog Remote Control of Voltage and Current Settings].
The output current can be controlled by external voltage.
(0 to rated voltage at the input voltage 0 to 10Vdc)
6-7 Monitoring Output Voltage/Current
Output voltage current value can be monitored by 0 to 10V monitor voltage output.
(The output voltage comes to 0 to 10 Vdc at 0 to rated voltage.)
(The output current comes to 0 to 10 Vdc at 0 to rated current.)
The recommended input impedance for connected measurement devices, etc. is at least 1 MΩ.
N
ote that when items with input impedance of around 10 kΩ are used, the monitor output voltage
value drops by approximately 10% on the measurement device.
Input impedance 10 k
Ω
→
Output monitor voltage drops by approximately 10%
100 k
Ω
→
Output monitor voltage drops approximately 1%
Fluctuation of external control voltage source will directly affect the boutput voltage. Prepare stable
voltage source.
TB1
0 to 10Vdc
External connection
+
2
5
13
25
14
1
Internal circuit
of unit
Common
Internal circuit
of unit
Common
0 to 10Vdc
-
+
1kΩ
0 to 10Vdc
-
+
1kΩ
TB1
Output current
monitor
0 to 10Vdc
External connection
3
5
13
25
14
1
15
- +
V
- +
V
Output voltage
monitor
0 to 10Vdc
Содержание HARS Series
Страница 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Страница 10: ......
Страница 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Страница 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Страница 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Страница 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Страница 92: ......