3 Connecting
17
HARS series
3-3 Connecting the load
Turn off the POWER ON/OFF switch.
(For safety, it is recommended that the AC input cable be disconnected during the operation.)
Be sure to ground the main unit as well as the load frame.
Follow the connection above to connect the shielded wire with the supplied high voltage cable.
After turing off the high voltage output, note the following points when you touch the load:
1. Set the output voltage setting to zero, or set the [OUTPUT ON/OFF] switch to OFF.
2. Confirm that the voltage is zero on the output voltmeter of the product.
3. Turn OFF the POWER ON/OFF switch.
4. Ground it more than 10 seconds. Then, check the voltage with a high-voltage
voltmeter.
It is especially dangerous
if the load is capacitive or if a cable longer than the
supplied cable is used
.
5. Use the right hand when operating the load.
Содержание HARS Series
Страница 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Страница 10: ......
Страница 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Страница 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Страница 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Страница 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Страница 92: ......