5 Operation on the Main Unit
29
HARS series
Arc count (2)
5-1-9-2
The output is turned off when a set number of arcs have occurred within a set time after the arc has
occurred.
The arc count shows the number of times the arc has occurred within the set time. By touching the
Arc count reset button, the arc count can be reset to "0".
To enable the function of turning off the output according to the arc count, touch the Arc count button
and set it to "Enable".
When the arc protect count setting value is touched, the protect count value for the arc for which the
output is to be turned off can be set with the setting change dial.
Touch the "Arc protection enable time" to set the enable time of the arc count to turn off the output
with the setting change dial.
When the arc count exceeds the arc protect count setting value, the output is turned off.
Setting screen
Home screen
Arc count
Arc count button
Arc total count
Arc count reset button
Setting value of Arc protect
count
Arc protection enable time
Arc count (2) setting screen