5 Operation on the Main Unit
45
HARS series
On the Ramp setting screen, touch the Ramp ON/OFF button to set it to Enable.
Set the ramp ON time (ON time). (0.0s to 99h59m59.9s)
Set the ramp OFF time (OFF time). (0.0s to 99h59m59.9s)
Sets the ramp mode. There are three modes to select for ramp operation: both voltage and current, voltage
only, and current only.
To use the multiset memory together, set the multiset function to Enable on the multiset setting screen.
Set the ramp time (A-ON time/B-ON time/C-ON time) for each of the multiset memory A/B/C.
(0.0s to 99h59m59.9s)
When multiset memory is enabled, the ramp ON time (ON time) is not used.
To change each setting value, touch each item on the touch panel display.
Example: In combination with the sequence function
Ramp
operation
in ON
OUTPUT
ON
1 cycle
OUTPUT
OFF
On time
Ramp
operation
in OFF
OFF time
Example: In combination with Pulse sequence function and Multiset memory
Ra
Output ramp time of Memory A
Ta
Memory A output time
Rb
Output ramp time of Memory B
Tb
Memory B output time
Rc
Output ramp time of Memory C
Tc
Memory C output time
Roff
Off ramp time
Toff
OFF time
OUTPUT
ON
1 cycle
Stop after
repeating
twice
Memory A
Memory B
Memory C
Output 0
Ta
Tb
Tc
Toff
Ra
Rb
Rc
Roff
Содержание HARS Series
Страница 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Страница 10: ......
Страница 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Страница 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Страница 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Страница 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Страница 92: ......