6 Operating by Analog Remote Control
51
HARS series
2) Set the current setting to Remote mode.
Home screen
When the voltage setting is set to Remote mode, "EXTV" will appear in the upper left corner of the
home screen.
6-5 Output Voltage Program by External Voltage
Confirm that it is set to analog remote control.
(See [6-4 Analog Remote Control of Voltage and Current Settings].
The output voltage can be controlled by external voltage.
(0 to rated voltage at the input voltage 0 to 10Vdc)
Fluctuation of external control voltage source will directly affect the boutput voltage. Prepare stable
voltage source.
TB1
TTL signal
External relay
External connection
+
-
25
17
13
25
14
1
Internal circuit
of unit
470Ω
TB1
0 to 10Vdc
External connection
+
1
5
13
25
14
1
Internal circuit
of unit
Common