Root
Chapter 5: NIBP Measurement
www.masimo.com
91
Masimo
3.
Properly place the blood pressure cuff on patient. See
Cuff Selection and
Placement
on page 85.
4.
To enable
Automatic
mode, touch the
Main Menu
icon
, then select
NIBP
Settings
.
On the
Intervals
screen, change
Set Mode
to
Automatic
, and then select the
desired
Interval
. The measurement mode can also be changed using the action
menu.
WARNING
: Too frequent blood pressure measurements can cause injury to the
patient due to blood flow interference.
5.
To begin measurement, touch the
Start Auto
button and then press the arrow at
the top-left corner of the touchscreen to return to the
Main View
.
6.
Wait for measurement to complete or touch the
Stop
button to stop the
measurement.