Nurse Call Setting Connections
For maximum flexibility, either normally open or normally closed signals are available. During
an alarm condition or a low Signal IQ event, depending on the configuration of the device
output, the normally open pin will be connected to the common pin and the normally closed
pin will be disconnected. In addition, the Nurse Call Polarity can be inverted to accommodate
various nurse call station requirements.
Only qualified personnel should connect one of these two signals to a hospital’s Nurse Call
system.
Cable Description
Nurse Call Event
Menu Setting
2 contacts normally opened
Nurse Call
Polarity Normal
2 contacts normally closed
Nurse Call
Polarity Inverse
1 and 2 contacts normally opened
2 and 3 contacts normally closed
Nurse Call
Polarity Normal
1 and 2 contacts normally closed
2 and 3 contacts normally opened
Nurse Call
Polarity Inverse
1 and 2 contacts normally closed
2 and 3 contacts normally opened
Nurse Call
Polarity Inverse
Battery Test
To conduct a Battery Test
1.
Fully charge Root by connecting it to AC power.
2.
Verify that the Root Charging Indicator is illuminated.
3.
When Root is fully charged, the Root Charging Indicator turns off.
4.
Power On Root and verify that the Root Battery Indicator icon on the Status Bar
shows a full charge.
Содержание Root
Страница 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 20: ......
Страница 80: ......
Страница 86: ......
Страница 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Страница 104: ......
Страница 106: ......
Страница 108: ......
Страница 112: ......
Страница 116: ......
Страница 136: ......
Страница 146: ......
Страница 150: ......
Страница 151: ......
Страница 152: ...www masimo com 38280 9274A 0316 ...