Root
Chapter 4: Temperature Measurement
www.masimo.com
82
Masimo
Temperature Probes
Two types of probes are available for use with the Root: an oral/axillary probe and a rectal
probe.
Root will automatically detect the probe type when connected: oral/axillary or rectal.
•
Adult axillary Spot Check Mode (predictive) temperatures are measured using the
oral/axillary probe in combination with the Root in Adult axillary mode. A
temperature reading is provided in approximately 12-15 seconds.
•
Pediatric axillary Spot Check Mode (predictive) temperatures are measured using
oral/axillary probes in combination with the Root in Pediatric axillary mode. A
temperature reading is provided in approximately 10-13 seconds.
•
For oral temperatures, place the probe tip under the patient's tongue on
either side of the mouth to reach the sublingual pocket and ask the patient
to close his/her lips.
Sublingual Pocket Location