Chapter 12: Specifications
www.masimo.com
121
Masimo
Alarms
Audio Alarm
Type
System Status
Light Color
Audio Description
High Priority Flashing Red
10-pulse burst, pulse spacing: 0.250s, 0.250s,
0.500s, 0.250s, repeat time:10s
Medium
Priority
Flashing Yellow
3-pulse burst, pulse spacing: 0.375s, 0.375s, repeat
time: 7s
Low Priority
Yellow
1-pulse burst, repeat time: 5s
Nurse Call Specifications
The Nurse Call relays have the following electrical specifications per switch:
Parameter Specification
Max Voltage 36 VDC or 24 VAC peak