Safety Information, Warnings, and Cautions
www.masimo.com
13
Masimo
WARNING
: Neonatal blood pressure measurements must always use a 3 meter hose in order
to avoid overpressure error caused by lack of air volume within the overall pneumatic system.
WARNING
: Before use, verify the color of the removable probe well to confirm the proper
application site: Red (rectal), Blue (oral/axillary).
WARNING
: Do not apply the cuff to a limb that is on the same side of a mastectomy.
WARNING
: Do not use or stop blood pressure measurements if the patient appears to be
affected by the pressurization of the cuff due to a physical condition (i.e. pregnant,
pre-eclamptic, etc.)
WARNING
: Too frequent blood pressure measurements can cause injury to the patient due to
blood flow interference.
WARNING
: Do not attach the cuff to a limb being used for IV infusions or any other
intravascular access, therapy or an arterio-venous (A-V) shunt. The cuff inflation can
temporarily block blood flow, potentially causing harm to the patient.
CAUTION
: Applying the blood pressure cuff over a wound can cause further injury.
CAUTION
: A compressed or kinked connection hose may cause continuous cuff pressure
resulting in blood flow interference and potentially harmful injury to the patient.
CAUTION
: Do not place the Root where the controls can be changed by the patient.
CAUTION:
To ensure patient isolation, connect only Masimo devices that have been designed
for Root.
CAUTION:
Equipment intended to be connected to signal input/signal output ports should
comply with applicable electrical safety standards to further minimize the risk of electric
shock. Only devices that have been configured to operate with Root may function properly
when connected.
CAUTION:
Only use the AC power cable provided by Masimo. Using a different AC power cable
could cause damage to Root. Check the power cord and plug to ensure that it is intact and
undamaged.
CAUTION:
To avoid risk of electrical shock, this equipment must only be connected to a
supply mains with a protective earth connection. Do not under any circumstances remove the
grounding conductor from the power plug.
CAUTION:
Use a grounded outlet for proper equipment grounding. A hospital-grade outlet is
required.
CAUTION:
Do not place Root where the appliance inlet or the AC power plug cannot be
readily disconnected.
Note:
Disconnect the device from AC mains by removing the AC power cord connector from
the device inlet.
Note:
If there is any doubt about the integrity of the protective earth conductor arrangement,
operate Root on internal battery power until the AC power supply protective conductor is fully
functional.
Note:
Do not monitor more than a single patient at a time on Root.
Note:
It is recommended that Root is attached to an AC power source when it is not in use to
ensure that the battery remains fully charged.
Note:
For medical technologies that require AC power, the battery should be adequately
charged to ensure backup power in case of AC power disruption.
Содержание Root
Страница 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 20: ......
Страница 80: ......
Страница 86: ......
Страница 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Страница 104: ......
Страница 106: ......
Страница 108: ......
Страница 112: ......
Страница 116: ......
Страница 136: ......
Страница 146: ......
Страница 150: ......
Страница 151: ......
Страница 152: ...www masimo com 38280 9274A 0316 ...