Root
Chapter 3: Operation
www.masimo.com
77
Masimo
Radical-7 and Radius-7 Charging Indicator
When Root is connected to an AC power source, it is able to charge a correctly docked
Radical-7 or Radius-7. This is true whether the device is powered on, in Sleep Mode, or
powered off. Conversely, when Root is not connected to AC power, it will not charge the
device.
The light status provides a visual indication of the battery condition:
Light Status Battery Indication
Green
Battery is fully charged.
Orange
Battery is charging.
Red
Battery is unable to charge.
Off
Battery is fully charged, not being charged.
Sleep and Power Off
To put Root in the Sleep Mode or Power Off Mode, follow these steps:
State
Description
Sleep
Mode
Press and hold the Power Button for two (2) seconds until one (1) audible tone
sounds.
Sleep Mode conserves power while enabling a quicker startup sequence.
Power Off
Mode
Press and hold the Power Button for eight (8) seconds, until two (2) audible
tones sound. The Home Button will flash, and the Power Button will flash orange.
Power Off Mode completely shuts down Root and results in a longer startup
sequence.
Содержание Root
Страница 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 20: ......
Страница 80: ......
Страница 86: ......
Страница 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Страница 104: ......
Страница 106: ......
Страница 108: ......
Страница 112: ......
Страница 116: ......
Страница 136: ......
Страница 146: ......
Страница 150: ......
Страница 151: ......
Страница 152: ...www masimo com 38280 9274A 0316 ...