Root
Chapter 3: Operation
www.masimo.com
65
Masimo
Pulse Rate (PR)
From the
Pulse Rate
screen, touch
Alarms
, and then change any of the following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
Configurable Options
High
Limit
The High Limit is upper
threshold that triggers an
alarm.
High
120
40-215 in steps of 5,
or Off
When set to Off,
alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
High
50
35-210 in steps of 5,
or Off
When set to Off,
alarm is disabled