Chapter 5: NIBP Measurement
Root provides noninvasive blood pressure readings through an oscillometric method.
An oscillometric method of blood pressure measurement is a noninvasive method that
monitors the amplitude of cuff pressure changes during cuff deflation to determine arterial
blood pressure. The NIBP module is designed for use on adult, pediatric, and neonatal
patients.
Patient Measurement Mode
Below is a table that provides a method for selecting the appropriate NIBP patient category.
To change patient category see
Profiles
on page 53.
Weight
Patient Category Maximum Pressure
Greater than 75lbs (34kg)
Adult
280 mmHg
Between 15.4-75lbs (7-34kg) Pediatric
280 mmHg
Less than 15.4lbs (7kg)
Neonatal
140 mmHg
Cuff Selection and Placement
Root uses a quick connect hose with a blood pressure cuff to measure NIBP. Based on the Cuff
Type that is selected, use the following approved blood pressure cuff chart to choose sizes.
Accessory
Length
NIBP Patient Hose, Male Quick Connect 3 meters (9.8 feet)
NIBP Patient Hose, Female Bayonet
3 meters (9.8 feet)
Cuff Type
Size (color), Range - cm
Reusable Masimo Blood Pressure Cuff
Child (green), 12 - 19
Small Adult (turquoise), 17 - 25
Adult (navy blue), 23 - 33
Large Adult (burgundy), 31 - 40
Содержание Root
Страница 1: ...Operator s Manual Root with noninvasive blood pressure and temperature ...
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 20: ......
Страница 80: ......
Страница 86: ......
Страница 101: ...Root Chapter 6 Admit and Discharge to Patient SafetyNet www masimo com 99 Masimo Fig 6 ...
Страница 104: ......
Страница 106: ......
Страница 108: ......
Страница 112: ......
Страница 116: ......
Страница 136: ......
Страница 146: ......
Страница 150: ......
Страница 151: ......
Страница 152: ...www masimo com 38280 9274A 0316 ...