40
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
4.10.2 Disconnect Phase
The system automatically enters the
disconnect phase when the patient is
disconnected during the preparation phase.
During the disconnect phase the following
alarms are turned off for up to 60 seconds:
• Apnea
• Minute volume
• Frequency alarm
• EtCO
2
• PEEP
When the patient is reconnected, the system
automatically enters the post-oxygen phase
and restarts ventilation.
It is also possible to restart the ventilation
manually:
1. Press the
Start ventilation
touchpad to
restart ventilation manually.
4.10.3 Post-Oxygen Phase
After reconnection, the ventilator will
deliver the same oxygen concentration as in
the preparation phase for 60 seconds.
After 60 seconds the system automatically
returns to ventilation using the previous
oxygen setting.
4.11 Recalibrating the
Oxygen Cell
If the ventilator has been in continuous use
for an extended period, the measured O
2
concentration may drop due to normal
degradation of the oxygen cell. In order to
avoid nuisance alarms in this situation, it is
possible to temporarily adjust the O
2
cell
during ventilation.
When you activate the
O
2
cell adaptation
function, the oxygen cell is recalibrated so
that the current measured O
2
concentration
is equal to the O
2
concentration set by the
user. This temporary adjustment will be
valid until the ventilator is switched off.
Important:
Before using the Servo-i, always
perform a Pre-Use Check to make sure the
O
2
cell is properly calibrated.
To recalibrate the O
2
cell:
1. Press the fixed key
Menu.
2. Press the
Biomed
touchpad.
3. Press the
O
2
cell adaptation
touchpad.
4. Press the
Yes
touchpad to perform the
O
2
cell adaptation.
SVX-9017_XX
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...