31
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
4.4 Start-up Configuration
The ventilator will always start up with the
previous Start-Up Configuration. The Start-
Up Configuration can be edited, copied, and
saved.
You can edit the following Start-Up
Configuration settings:
• Patient category (
Adult
or
Infant
)
• Type of ventilation (
Invasive
or
NIV
)
• Volume setting
• Breath cycle setting
• Pre/post oxygenation concentration above
set O
2
concentration (%)
• Mode of ventilation (including parameter
settings)
This section provides instructions for
accessing and editing the start-up
configuration, changing the patient
category, and changing the type of
ventilation.
4.4.1 Accessing the Start-Up
Configuration
Note:
The ventilator must be in Standby
mode.
1. Press the fixed key
Menu.
2. Press the
Biomed
pad and enter the
access code (the factory setting is
1973).
The
Biomed
submenu consists of the
following touchpads:
• Service
• Edit configuration
•
Copy configuratio
n
• Set date and clock
• Change access code
To alter ventilator settings, press the
appropriate touchpad and follow on-screen
instructions.
4.4.2 Editing the Start-up
Configuration
To edit the Start-up Configuration.
1. Press the
Edit configuration
touchpad in
the
Biomed
submenu.
2. Press the
Start-up configuration
touchpad.
3. Press the touchpad for desired start-up
setting.
4. Press
Next
to continue to ventilation
mode settings.
5. Press the appropriate touchpad to
change the settings.
Note:
Press
Restore mode settings
to restore
factory default settings.
6. Press
Next
to view a summary of the
start-up configuration
7. Press
Accept
to save the start-up
settings.
Note:
The ventilator must be restarted to
activate the new settings.
SVX-9046
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...