89
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
10.5 Expiratory channel
Pressure Drop
Maximum: 3 cmH
2
O at a flow of 1 liter/s
Internal Compressible Factor
Maximum: 0.1 ml/cmH
2
O
PEEP Regulation
Microprocessor controlled valve
PEEP Setting Range:
0 - 50 cmH
2
O ± 5% or ± 1 cmH
2
O
1
Expiratory Flow Measurements
Flow range: 0 - 3.2 liters/s ±5% or ± 2.5 ml/s
Rise time (flow of 0.05 - 3.21 liters/s): < 12 ms for 10 -
90% response ±5% or ± 2.5 ml/s
10.6 Monitoring
Expiratory Minute Volume
Adult Range: 0 - 60 liters/min
± 8% or ± 0.15 liters/min
2
, NIV: ± 10%
3
Infant Range: 0 - 20 liters/min
± 8% or ± 0.15 l/min
4
, NIV: ± 10%
5
, NIV Nasal CPAP: ±
25% or ± 0.15 liters/min
6
Expiratory Tidal Volume
Adult Range: 0 - 2000/4000 ml ± 8% or ± 18 ml
7
Infant Range: 0 - 350 ml ± 8% or ± 2 ml
8
O
2
Concentration
Range: 0 - 100% ± 5% of read value
Airway Pressure
Range: -40 - 160 cmH
2
O ±5% or ± 1 cmH
2
O
Supply Pressure
Range: 0 - 7 bar ± 5% of read value
10.7 Alarms
10.7.1 Allowed Alarm Settings
Airway Pressure (upper)
Adult: 16 - 120 cmH
2
O
Infant: 16 - 90 cmH
2
O
High Continuous Pressure (> 15 sec)
Set PEEP level+15 cmH
2
O
O
2
Concentration
Set value±6vol% or < 18 vol.%
Expired Minute Volume (Upper alarm limit)
Adult: 0.5 – 60 liters/min
Infant: 0.01 – 30 liters/min
Expired Minute Volume
(Lower alarm limit)
Adult: 0.5 – 40 liters/min
Infant: 0.01 – 20 liters/min
Apnea
Adult: 15 - 45 s
Infant: 5 - 45 s
9
Gas Supply
< 2.0 kPa x 100 or > 6.5 kPa x 100.
Respiratory Frequency
1 – 160 b/min.
High End Expiratory Pressure
0 - 55 cmH
2
O
Low End Expiratory Pressure
0 - 47 cmH
2
O
10
End-Tidal CO
2
(upper and lower limit
11
)
0.5-20%, 4-100 mmHg, 0.5-14 kPa
CPAP (Upper alarm limit)
Adult: 0-55 cmH
2
O
Infant: 0-55 cmH
2
O
CPAP (Lower alarm limit)
Adult: 0-47 cmH
2
O
Infant: 0-47 cmH
2
O
1. at RR < 60 b/min
2. at RR < 100 b/min
3. at constant leakage fraction <30%
4. at RR < 100 b/min
5. at constant leakage fraction <30%
6. at constant leakage fraction <30%
7. at Expiration time < 4 s and RR <100 b/min
8. at Expiration time < 1 s and RR <100 b/min
9. The apnea alarm can be turned off in Nasal
CPAP.
10.
Note:
Setting the alarm to 0 (zero) is equal to
alarm off.
11. In NIV low limit can be set to 0 (zero).
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...