![Maquet SERVO-i Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/maquet/servo-i/servo-i_user-manual_1724263080.webp)
30
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
4.3 Patient Circuit Test
In Standby mode, the Patient Circuit Test
may be performed separately from the Pre-
Use Check. This is useful, for example, when
changes are made to the circuit or additional
accessories are connected. The test
evaluates circuit leakage and measures the
circuit compliance.
1. Press the
Patient circuit test
touch pad
and follow the on-screen instructions.
WARNINGS!
• A Pre-Use Check must always be done
before connecting the ventilator to a
patient.
• The Patient Circuit Test does not replace
the Pre-Use Check.
Note:
Considerable leakage may occur
around the endotracheal tube if it is
uncuffed. The combination of small tidal
volumes, leakage around the tube, and
activated compliance compensation may
trigger the
Low Expiratory Minute Volume
alarm due to a very low expiratory flow
passing from the patient through the
expiratory channel. By observing the
difference between the Vti and Vte values
presented on the user interface, a leakage
can be detected and its extent easily
controlled. The first time an unacceptably
large leakage occurs around the tube,
correct this problem to avoid triggering the
Low Expiratory Minute Volume
alarm. If the
leakage persists, adjust the alarm limit
down to its lowest level (10 ml)—if this step
is clinically appropriate. Finally, if the
leakage still has not been remedied,
deactivate the compliance compensation to
avoid triggering the alarm. If the compliance
compensation is deactivated while in
Pressure Control, Pressure Support, or SIMV
(Pressure Control) ventilation modes, then
no further settings need to be adjusted.
However, in volume-related modes, the set
volumes must be adjusted.
SVX-9042
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...