55
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
6.1.3 Summary
Ventilation Mode
Abbreviation
Notes
a
Pressure Regulated
Volume Control
PRVC
ALARM if target volume cannot be delivered due to
setting of UPL; cm; eb; df
Volume Control
VC
IMPORTANT to set UPL; cm; eb; cf
Pressure Control
PC
IMPORTANT to set alarm limits for MVol; cm; eb; df
Volume Support
VS
ALARM if target volume cannot be delivered due to
setting of UPL; IMPORTANT to set alarm limits for MVol,
Apnea; pib
Pressure Support
PS
IMPORTANT to increase inspiratory rise time for
patient comfort, set alarm limits for MVol, monitor
TVol; pib
Continuous Positive
Airway Pressure
CPAP
IMPORTANT to set alarm limits for MVol, Apnea; sb
Synchronized
Intermittent Mandatory
Ventilation (PRVC) + PS
SIMV (PRVC) +
PS
comb. control and pressure support/spontaneous
function
SIMV (VC) + PS
SIMV (VC) + PS
comb. control and pressure support/spontaneous
function
SIMV (PC) + PS
SIMV (PC) + PS
comb. control and pressure support/spontaneous
function
Bi-Vent
BV
IMPORTANT to set alarm limits for MVol, Auto Scale not
recommended, Apnea alarm and back-up ventilation
not available, two pressure levels and durations are
set, alarms are handled for each Bi-Vent cycle
Non-Invasive
Ventilation - PC
NIV-PC
for ALL NIV modes: avoid high inspiratory pressure, use
of nebulizer not recommended, excessive leakage will
cause a high-priority ALARM, trigger sensitivity and
cycle off cannot be set; NIV - PC is a controlled
breathing mode (cm)
NIV - PS
NIV-PS
pib
NIV - nasal CPAP
Nasal CPAP
IMPORTANT to have a ventilator-independent means of
monitoring in place, no backup ventilation available,
Apnea alarm can be turned off; sb
Automode
none
VC <--> VS or PRVC <--> VS or PC <--> PS; not in NIV
a. MVol = Minute Volume, UPL = Upper Pressure Limit, TVol = Tidal Volume, cm = controlled ventilation mode, pib
= patient-initiated breathing mode, sb = spontaneous breathing mode, eb = patient can trigger extra breaths,
df = decelerating inspiratory flow, cf = constant inspiratory flow
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...