9.3 Medium Priority Alarms
Alarm message
Possible causes
Remedies
Air supply pressure:
High
Air supply pressure above 6.5 kPa x 100.
Air supply pressure at gas inlet is too
high.
Check the gas supply lines. Perform
a Pre-Use Check.
Contact a service technician.
Air supply pressure:
Low
Air supply pressure below 2.0 kPa x 100.
Air supply pressure at gas inlet is too
low.
Gas supply line disconnected.
Note:
This alarm can be permanently
silenced (Audio off) when activated.
Check and connect gas supply
lines.
Perform a Pre-Use Check.
Alarm output
connection error
Technical problems (hardware or
software) with the external alarm
function.
Contact a service technician.
Battery mode!
Nebulizer switched
off
Ventilator is running on batteries and
the Servo Ultra Nebulizer is disabled to
reduce power consumption.
Connect to AC power to use the
Servo Ultra Nebulize.
Check the connection to AC power.
Battery operation
AC power interrupted.
Check the connection to AC power.
Check alarm limits
The persistent memory has corrupt
contents.
Check the alarm limits.
Check CO
2
airway
adapter
The data, reference channel, or both
are out of range. Usually occurs if the
airway adapter is removed from the
capnostat, there is optical blockage on
the windows of the airway adapter, or
the adapter type was changed and
verification was not performed.
Make sure that the adapter is
completely inserted. Clean airway
adapter if necessary.
Open the CO
2
Calibration Window
and perform Verification.
Check default alarm
limits
Problems in internal memory for default
alarm limits.
Check default alarm limits.
Contact a service technician.
Check Y Sensor
Y Sensor is not connected to the patient
breathing system or Y Sensor is not
working properly.
Check sensor connection to patient
breathing system.
Change the Y Sensor.
CO
2
module error
Hardware error in the CO
2
Analyzer
module.
Unplug and reinsert the module.
Change the module.
Contact a service technician.
CO
2
module
unplugged
CO
2
Analyzer module is not properly
inserted.
Insert the CO
2
Analyzer module.
CO
2
sensor
disconnected
CO
2
capnostat sensor is not attached.
Connect the sensor to the CO
2
Analyzer module.
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...