Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
9.7
CO
2
Analyzer: Calibration Error Messages
Message
Description
Remedies
Adapter zero: Failed
An error was detected during the
verification calibration (adapter
zero). The airway adapter is
occluded or CO
2
gas is present in
the adapter. The calibration was
cancelled and old cell zero
parameters were not restored.
Perform a verification calibration.
Hardware problem: contact a
service technician.
CO
2
cell zero failed
An error was found during cell zero
calibration. The calibration was
cancelled and old cell zero
parameters were not restored.
Perform a cell zero calibration.
Hardware problem: contact a
service technician.
Verification against
Reference cell: Failed
The capnostat sensor is faulty or
there is an optical blockage of the
capnostat sensor windows.
Clean the capnostat sensor
windows.
Replace the capnostat sensor.
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...