90
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
10.7.2 Autoset Alarm Limits
High Airway Pressure:
Mean peak pr10 cmH2O or at least 35 cmH2O.
Expiratory Minute Volume (upper)
+ 50%
Expiratory Minute Volume (lower)
- 50%
Respiratory Frequency (upper)
+ 40%
Respiratory Frequency (lower)
- 40%
Mean End Expiratory Pressure (high)
+5 cmH
2
O
Mean End Expiratory Pressure (low)
-3 cmH
2
O.
etCO
2
Concentration (upper)
+ 25%
etCO
2
Concentration (lower)
- 25%
10.7.3 Alarms Miscellaneous
Audio Pause (Alarm silence/reset)
Two-minute silence and reset of latched alarms.
10.8 Ventilation modes
10.8.1 Controlled Ventilation
Pressure Control (PC)
Pressure controlled ventilation.
Volume Control (VC)
Volume controlled ventilation.
Pressure Regulated Volume Control (PRVC)
Pressure regulated volume controlled ventilation.
NIV Pressure Control
Noninvasive pressure controlled ventilation.
10.8.2 Supported ventilation
Volume Support (VS)
Volume supported ventilation.
Pressure Support (PS)/CPAP
Pressure supported ventilation / Continuous positive
airway pressure ventilation.
NIV Pressure Support
Noninvasive pressure supported ventilation.
Nasal CPAP
Nasal continuous positive airway pressure ventilation.
10.8.3 Combined ventilation
SIMV (VC) + PS
Synchronized intermittent mandatory ventilation based
on volume controlled ventilation with pressure support.
SIMV (PC) + PS
Synchronized intermittent mandatory ventilation based
on pressure controlled ventilation with pressure
support.
SIMV (PRVC) + PS
Synchronized intermittent mandatory ventilation based
on pressure regulated volume controlled ventilation
with pressure support.
Bi-Vent
Pressure controlled ventilation that allows the patient
the opportunity of unrestricted spontaneous breathing.
Automode
Control mode
Support mode
VC
<----------- ----------> VS
PC
<----------- ----------> PS
PRVC <----------- ----------> VS
In Servo-i flow measurements all preset and indicated
volumes are referenced to ambient pressure at +21˚C
(AP21).
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...