![Maquet SERVO-i Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/maquet/servo-i/servo-i_user-manual_1724263089.webp)
39
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
4.10 Using Suction Support
The Suction Support function makes it
possible to automatically inhibit the
ventilator from cycling during a tracheal
suction procedure without activating
alarms.
Suction Support includes:
• preparation phase
• disconnect phase
• post-oxygen phase
WARNING!
Suction Support is not intended
to be used together with closed-suction
systems.
WARNING!
The minimum PEEP level during
suction support is 3 cmH
2
O. The ventilator
will adjust to minimum level if the PEEP
level is below 3 cmH
2
O in order to detect
disconnection of the patient.
Important:
Alarms are turned off during the
disconnect phase for a maximum of 60
seconds. If the patient has not been
reconnected within 60 seconds, all alarms
are activated.
Note:
Suction Support is not available in NIV
mode or when the
O
2
Breaths
function is
activated.
Note:
During the disconnect phase in
Suction Support, the nebulizer is
temporarily paused.
Note:
When only one gas is connected, an
elevated oxygen level cannot be set during
the preparation phase. In this case, the post-
oxygen phase will be skipped.
4.10.1 Preparation Phase
To enter the preparation phase:
1. Press the fixed key
Quick access.
2. Press the
Suction Support
touchpad.
3. Set the desired pre-oxygen value by
turning and pressing the Main Rotary
Dial.
The following alarms are turned off:
•
Check tubing
The maximum duration of the preparation
phase is 120 seconds. After 120 seconds, the
system automatically returns to ventilation
using the previous oxygen setting.
Note:
The
Cancel
pad will close the Suction
Support program.
SVX-9000 XX
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...