![Maquet SERVO-i Скачать руководство пользователя страница 116](http://html1.mh-extra.com/html/maquet/servo-i/servo-i_user-manual_1724263116.webp)
66
Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
7.3.5 Alarm Ranges and Defaults
Alarm (priority)
a
Allowed Range; (Factory Default Value)
Audio Off
O
2
concentration high (HP)
NA; (Set Value+6vol%)
No
O
2
concentration low (HP)
NA; (Set Value-6vol% or
≤
18vol%)
No
High continuous pressure (HP)
NA; (Set PEEP level+15 cmH
2
O for > 15 sec)
No
P
aw
high (HP)
b
Adult: 16 - 120 cmH
2
O; (40, 20 in NIV)
Infant: 16 - 90 cmH
2
O; (40, 20 in NIV)
No
Apnea (HP)
Adult: 15 - 45 sec;(20)
Infant: 5 - 45 sec; (10)
No
Expired minute volume high
(HP)
Adult: 0.5 - 60 l/min; (40)
Infant: 0.01 - 30 l/min; (5)
Yes
Expired minute volume low
(HP)
Adult: 0.5 - 40 l/min; (5)
Infant: 0.01 - 20 l/min; (2)
Yes
Respiratory rate high (MP)
Adult: 1 - 160 b/min; (30)
Infant: 1 - 160 b/min; (50)
Yes
Respiratory rate low (MP)
Adult: 1 - 160 b/min; (5)
Infant: 1 - 160 b/min; (20)
Yes
PEEP high (MP)
0 - 55 cmH
2
O; (10)
Yes
PEEP low (MP)
c
0 - 47 cmH
2
O; (10)
Yes
etCO
2
high (MP)
d
0.5 - 20%; (6.5)
4 - 100 mmHg; (49)
0.5 - 14kPa; (6.5)
Yes
etCO
2
low (MP)
0.5 - 20%; (4)
4 - 100 mmHg; (30)
0.5 - 14kPa (4)
Yes
CPAP high (MP)
Adult: 0 - 55 cmH
2
O; (10)
Infant: 0 - 55 cmH
2
O; (10)
Yes
CPAP low (MP)
Adult: 0 - 47 cmH
2
O; (10)
Infant: 0 - 47 cmH
2
O; (10)
Yes
a. HP = High priority alarm, MP = Medium priority alarm
b. If P
aw
rises 6 cmH
2
0 above the set limit or if system pressure exceeds 117 ± 7 cmH
2
0, the safety valve opens.
c. Setting the alarm limit to 0 (zero) is equivalent to turning the alarm off.
d. If the alarm limit is set outside the measuring range, no alarm will be activated even if the limit is exceeded.
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...