Servo… User´s manual
English edition
Order No: 65 14 892 E313E
Infant
Adult
Universal
Options
10.12 Immediate functions
Oxygen Breaths
100% for 1 minute.
Start Breath
Initiation of 1 breath in all modes.
(In SIMV mode initiation of one mandatory breath).
Pause Hold
Inspiratory or expiratory.
10.13 Communication/
Interface
Serial Port
RS-232C-isolated. For data communication via the
Communication Interface Emulator (CIE).
Alarm Output Connection Option
Isolated 4-pole modular connector for communication
of high and medium priority alarms. The alarm output
connection option is a non-guaranteed alarm in
accordance with IEC60601-1-8. Max 40 V DC, Max 500
mA, Max 20 W.
Data Transfer via Ventilation Record Card
File format is Unicode (big endian). Required software
is Microsoft Excel 2000 with Visual Basic for
applications.
Service
• Preventive maintenance must be performed by
authorized personnel at least once every year as
long as the unit is not operated for more than 5000
hours per year. The
Status
menu on the user
interface shows the current operating time.
• Battery modules must be replaced every 3 years.
• Original parts from MAQUET must be used.
• Service repair must be performed by MAQUET-
authorized personnel only.
• Service mode should only be used without a patient
connected to the ventilator.
10.14 Servo Ultra Nebulizer
Patient Unit
Weight, approximate (g):
125
Dimensions (mm):
60 W x 108 L x 105 H
Nebulizer T-Piece Connections
Inlet/outlet:
22/15 mm outside/inside diameter
22 mm inside diameter, ISO standard
Infant patient tubing, nipple connectors:
22/10 mm outside diameter
15/10 mm outside diameter
Internal Volume (ml)
60
Ultrasonic Generator Frequency (MHz)
2.4
Particle Size, approximate (water, µm)
4.0, mass median diameter (MMD) measured distally in
endotracheal tube 8 mm inside diameter.
Output from nebulizer (water)
Minimum water flux:
0.1 ml/min at gas flow 0.1l/s
0.3 ml/min at gas flow 0.5l/s
0.5 ml/min at gas flow 1.0l/s.
Buffer Liquid
Sterile water
Maximum Medication Temperature
55º C (131º F)
Volume, medication cup (Maximum, ml)
10
Noise level (Maximum, dBA)
50, measured at 0.3 m distance
Connection Cable Length (m)
2.0
Note:
For information about the stand-alone Aeroneb
Professional Nebulizer System, refer to accompanying
documentation.
Содержание SERVO-i
Страница 2: ...2 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA VENTILATION PARAMETERS...
Страница 10: ...10 MCV00020395 REVA VENTILATOR PARAMETERS MCV00020395 REVA MCV00020395 REVA CATEGORIES OF VENTILATION MODES...
Страница 41: ...41 MCV00020395 REVA MCV00020395 REVA COMBINATION MODES APRV BiLevel BiPhasic DuoPAP Similar names for Bi Vent...
Страница 43: ...User s manual VENTILATOR SYSTEM SERVO i V3 0...
Страница 44: ......
Страница 50: ...vi...
Страница 68: ...18 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 92: ...42 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 102: ...52 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...
Страница 110: ...60 Servo User s manual English edition Order No 65 14 892 E313E Infant Adult Universal Options...