Appendix A
Appendices
Page A-4
MAN 0318
Site requirements
MPT-2 Multi-purpose titrator
Note
These requirements should be read in conjunction with the system
requirements as stated in the Zetasizer Nano user manual.
Nitrogen purge specification
Warning!
A Nitrogen supply must be used in a well ventilated environment.
Turn
off
the supply when not in use.
If a nitrogen supply is required then it must conform to the following specifications
and flow rate conditions:
It is important that the nitrogen supply is dry, free from oil and filtered to
remove any contaminants that could affect the sample being measured.
The flow rate should be adjustable between 2 and 20ml/min.
Vacuum degasser
Space Requirements
The Vacuum Degasser is designed to sit on a standard laboratory bench top, and is
plumbed into the MPT-2 Titrator system between the titrant supply and titrator
pump.
Enough space should be given to allow easy access to all components of the system.
Allow additional space both in front, to accommodate the tubing connected to the
unit, and behind to accommodate the power lead.
Dimensions
Height: 5.0 in.
Width: 2.87 in.
Depth: 9.81 in (distance front to back)
MAN0318-4.0 - MPT-2 Autotitrator-Degasser.book Page 4 Tuesday, September 4, 2012 12:15 PM