Maintenance
Chapter 5
MPT-2 & Degasser
Page 5-7
tinuing. (After a pH calibration has finished,
pH probe output (mV)
will
change to
pH probe output from calibration (mV)
.
6.
When the calibration of the first pH point is complete, the pH point will show
a
alongside.
7.
Remove the probe from the first pH buffer, rinse with deionised water and dab
dry.
8.
Continue the calibration until all required pH points have been calibrated.
During the calibration a
progress bar
at the bottom of the dialogue indicates how
far the calibration routine has proceeded.
The ticks or crosses alongside the pH probe indicate the following:
Points can be deleted by selecting the point and pressing
Remove
.
Choosing the buffers when calibrating
The default buffers for calibrating are pH 4 and pH 9. Other values may be used
but ensure that the lowest and highest values are at least 3 pH values apart. Also if
all titration measurements tend to be in the lower end of the pH table it is advisable
to use pH 2 and pH 5 buffers to calibrate the probe.
Adding a pH buffer point
To use a different pH value, a new pH buffer point must be added to the selection
list. Press
Add
to display the
Calibration point
dialogue, then perform either of
the below actions:
1.
Select a defined pH point from the
Calibration pH
list and press
OK
to add it
to the
Calibration points
list in the main dialogue.
2.
Or if using a pH buffer not available in the list, select
Custom
, type the
value in the
Custom buffer
entry box, then press
OK
to add it to the
Cali-
bration points
list in the main dialogue.
If the maximum accuracy of calibration is required over a wide pH range, then 3 or
4 buffers should be used
The pH probe for the Titrator is supplied with sachets of pH4 and pH9 buffers.
Follow the instructions supplied with the sachets for the best methods of making
the buffer solutions.
A pH point has been chosen but not yet calibrated.
A pH point has been chosen and been calibrated.
Calibration points selected during a previous calibration, which have not
been calibrated during the current session. If required select these points
and perform the calibration process as described. A tick will appear once
it is complete.