Software and Control
Chapter 3
MPT-2 & Degasser
Page 3-17
pH calibration dialogue
With this dialogue the pH calibration points can be set and the calibration per-
formed.
It is important that the pH probe is correctly calibrated on a regular basis. If the
probe is not calibrated then the points of the measurement may be incorrect.
It is recommended that the calibration should be performed at the beginning of
each titration session.
The
pH calibration
dialogue is displayed by selecting
Tools-Instrument-
MPT-2 Titrator- pH calibration
within the Zetasizer Nano software, or by
selecting the
pH calibration
icon. Brief instructions on performing the cali-
bration appear on the dialogue.
Calibration points can be added or removed using the respective buttons. Adding a
pH point will display the
Calibration point
dialogue; input a pH point and press
Ok
to add it to the
calibration points
list.
Calibration can proceed by simply selecting the pH buffer points to use and press-
ing the
Measure
button. Guidance instructions on performing the calibration will
appear on the dialogue.
A
progress bar
shows how far the calibration routine has proceeded. A tick
is
displayed once a pH point has calibrated.