Connection and accessory preparation
Chapter 2
MPT-2 & Degasser
Page 2-11
Calibrating the pH probe.
The performance of the titrator relies on the pH probe being correctly calibrated. If
the pH probe has not been calibrated then the pH values will be incorrect.
It is recommended that the pH probe is calibrated before each titration session.
To calibrate the pH probe the follow the instructions in the
Software
and
Main-
tenance
chapters.
Preparing the sample in the container
Before starting the measurement, it is necessary to prepare the sample and attach
the container to the dispersion head
First prepare the sample to be measured and place a known volume into a sam-
ple container. An initial sample volume of 10ml is recommended (5ml mini-
mum to 25ml maximum). The preparation of the sample is of prime
importance. Read the sections on sample preparation in the Zetasizer Nano
manual.
Note
The sample used should be of a suitable concentration that gives an inten-
sity of scattering within recommended values. This is checked at the
beginning of each titration. If the concentration is too high the titration
may proceed slowly due to the sample buffering the pH. If a message
occurs indicating the sample concentration is higher than recommended,
dilute the sample and restart the titration.
Attach the sample container to the dispersion head as described in the
Intro-
duction and Hardware
chapter.
Filling the cell
With the cell tubing connected, and the sample container prepared and mounted
onto the dispersion head, all that is left to do before configuring the SOP and start-
ing the measurement is to fill the cell.
The cell must be filled so that no bubbles are left within the sample path; this is
especially important for Zeta potential measurements. Follow the instructions
below:
Open the
Manual control
dialogue by selecting
Tools-Instrument- MPT-2
Titrator- Manual control
within the Zetasizer Nano software, or by select-
ing the
Manual control
icon.