Chapter 3
Software and Control
Page 3-18
MAN 0318
Also displayed in the dialogue is the
pH probe output from calibration (mV)
value. This is a direct voltage reading generated by the pH calibration circuitry and
corresponds directly to the
pH point from calibration
value alongside. During
the calibration routine this value should constantly change. If it does not then the
probe is either not immersed sufficiently in the pH solution, or the probe is broken
or disconnected.
While a pH calibration is being performed
pH probe output from calibration
(mV)
will change to
pH probe output (mV)
, and
pH point from calibration
will change to
Estimated pH
.
To calibrate the pH probe the follow the instructions in the
Maintenance
chapter.