Maintenance
Chapter 5
MPT-2 & Degasser
Page 5-9
ill 8743
Replacement is the reverse of the above procedure.
1.
Push the o-ring into the recess on the manifold
.
2.
Slide the syringe mount up, checking that the syringes smoothly enter the
manifold assembly - ensure the syringe does not bend severely.
3.
Prime the system once all the syringes have been replaced.
Changing the filter and tubing
The filter removes all particles above the grading size of the filter fitted. To main-
tain optimum filtration it is recommended it should be replaced regularly. Replace-
ment of the filter and any tubing on the titrator is performed using very similar
techniques.
Generally, all internal tubing is connected with the screw-fit connections and all
external tubing and filters use a 'Luer' style. The diagram below shows how the
connections are made. Some connections differ slightly but the principle is the
same.
To replace the tubing on the Sizing flowcell, simply pull the used silicone tubing
off the harder PTFE tube and replace with new tubing. Connect the tubing at the
tubing connection bracket as shown below.
1
2
3