Maintenance
Chapter 5
MPT-2 & Degasser
Page 5-13
Yellow Status LED is on
steadily, pump is running
and RPM seems high.
Pump is in initial pull-
down phase or system's
degassing demand has
increased.
Typically normal opera-
tion, although if pump
speed continues to rise
for an extended period of
time (as heard by the
pitch of the stepper
motor) it could indicate a
potential fault condition.
Yellow Status LED is
flashing approximately 1
second off, 1 second on.
Vacuum pump is not run-
ning.
Possible system leak.
Contact your Service
Representative.
Yellow Status LED is
flashing approximately 2
seconds off, 1 second
on. Vacuum pump is not
running.
Possible sensor or Con-
trol Board fault.
Contact your Service
Representative.
Is there a way to check
whether the system is
operating correctly when
Power and Vacuum
green LEDs are illumi-
nated, but pump can't be
heard running?
Due to the design of the
pump and degasser, the
pump is virtually silent at
low RPM, even though
vacuum is good and
degassing is normal.
Place a hand on the top
of the unit. A slight
vibration can be felt indi-
cating the pump is oper-
ating at low RPM.
Bubbles appear through
the output tubing.
Loose fitting(s).
Tighten the input and
output fittings.
No titrant flow.
If a buffer titrant was left
in the degasser for some
time after use, it may
plug the degasser ele-
ments.
Use a different channel,
or connect the channel
to a beaker of the titrant
without the buffer. Draw
the titrant through the
channel to dissolve the
buffer. Do not push the
titrant through the chan-
nel. If this flushing action
does not work, contact
your Service Represent-
ative.
Problem
Probable Cause
Solution
MAN0318-4.0 - MPT-2 Autotitrator-Degasser.book Page 13 Tuesday, September 4, 2012 12:15 PM