Software and Control
Chapter 3
MPT-2 & Degasser
Page 3-3
The
Measurement type
options are:
Sample - General options
This dialogue includes information from
both
the Size or Zeta
Sample - General
options
dialogues. See description in the Zetasizer Nano manual for details. Only
the
Sample viscosity option
will be shown if an
Intensity only
measurement
option is chosen.
Sample - Temperature
See the description in the Zetasizer Nano manual.
Note
As the temperature probe is not temperature compensated, it is recom-
mend that titrations are performed with ambient temperature conditions.
Adjust the temperature in the SOP to the ambient value.
Titration - Titrants
This dialogue enables the titrants used to be specified, and the pump and stirrer
speed to be set.
Titration type
pH
This is a standard pH titration used to monitor solution prop-
erties, or sample size, as a function of sample pH. During a
Zeta potential pH titration the Isoelectric point can be deter-
mined.
Diluton
The Dilution titration is used to monitor either sample Size or
Zeta potential parameters as a function of sample concen-
tration.
Additive
This titration is used monitor the effects of formulation addi-
tives on the Size and Zeta potential of the sample. Results
are shown as a function of concentration.
Additive Log
(Conductivity)
This titration is used to monitor the conductivity of a sample
by adding salts. Results are shown as a function of conduc-
tivity
MAN0318-4.0 - MPT-2 Autotitrator-Degasser.book Page 3 Tuesday, September 4, 2012 12:15 PM