Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 92 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.7.14
:DGIO:SET:OUT
{x}
:FUNC[?]
Description:
Set or query the function of each output.
x is for outputs 1 to 4.
Mode:
Command or query
Parameter:
{
NO|IFHI|IFLO|IFIN|IFOU|IFOV|IFRN
}
Parameter
Function
NO
No function assigned to this output
IFHI
This output is set if the measuring value is geater
than the set upper limit value.
IFLO
This output is set if the measuring value is less than
the set lower limit value.
IFIN
This output is set if the measuring value is within the
range of upper and lower limit values.
IFOU
This output is set if the measuring value exceeds the
range of lower and upper limit values.
IFOV
This output is set if the measuring range has been
exceeded.
IFRN
This output is set if the device is ready to measure.
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:DGIO:SET:OUT2:FUNC IFOV<LF>
send
:DGIO:SET:OUT2:FUNC?<LF>
receive IFOV<CR><LF>
9.12.7.15
:DGIO:SET:OUT
{x}
:LOG[?]
Description:
Set or query the logic level of each output.
x is for outputs 1 to 4
Mode:
Command or query
Parameter:
{
0|1
}
0 is for normal function (current flow when this condition is fulfilled)
1 is for inverting function
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:DGIO:SET:OUT2:LOG 1<LF>
send
:DGIO:SET:OUT2:LOG?<LF>
receive 1<CR><LF>