Gaussmeter IGM11 Operating Instructions
Chapter 4 Control Elements and Connections
Page 20 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
4.3
Power Supply
The gaussmeter can be operated at a voltage of 12 V to 24 V DC. The current
consumption is approx. 200 mA at 12 V or approx. 100 mA at 24 V. An electrical
separation does not take place. Please notice that the shielding of the probes (if so,
the metal housing) will be connected with the earth conductor of your control via
this port.
4.4
Probe Connection
The measuring probe is plugged into the probe port (M8-socket IEC 61076-2-104 )
on the top of the device.
Note
Only use measuring probes which are approved by the manufacturer
to operate with this device.
Note
The probe plugs must not be connected to the electric potential, the
protective conductor or the plug shell. If you measure near current
conducting parts, make sure that there is sufficient distance and a
sufficiently good insulation.
Figure
8 Ports Bottom
For add-ons
Interface
Power Supply
LAN
Do not use
Port for PLC
signals
Port for power
supply for the
device and the PLC
signals
RJ45 port for the LAN
connection