Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 72 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.1.4
*ESR[?]
Description:
Standard
E
vent
S
tatus Enable Query
Reset or query the standard event release register. The gaussmeter
returns the bit pattern of the register as a decimal value.
Mode:
Command and query
Parameter:
{<Value>} within the range of 0..255
Response:
Standard event release register
*RST
Value:
not relevant
Example:
send
*ESR 22<LF>
send
*ESR?<LF>
receive 160<CR><LF>
9.12.1.5
*IDN?
Description:
Id
entificatio
n
Query
Query the gaussmeter identification character string. The gaussmeter
returns the following identification text:
MAGSYS-MAGNET-SYSTEME,IGM11,1,xxxxxx,C
Mode:
Query
Parameter:
none
Response:
Identification character string
*RST
Value:
not relevant
Example:
send
*IDN?<LF>
receive
MAGSYS-MAGNET-SYSTEME,IGM11,1,011111,C<CR><LF>
9.12.1.6
*OPC?
Description:
Op
eration
C
omplete Query
Return of "1" to the output buffer, after the command has been
executed. The command *OPC? can be placed at the end of a
command line in order to synchronize the controller with the
command processing. The command itself has no function.
Mode:
Query
Parameter:
none
Response:
1
*RST
Value:
not relevant
Example:
send
*OPC?<LF>
receive 1<CR><LF>
Trial Version
Subje
ct to chan
ge