Industrial Gaussmeter IGM11 Operating Instructions
Chapter 5 Operation
Page 29 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
The following shows the accuracies for DC-field measurements.
Field Strength
Accuracy (1
σ
)
≤ 1.5 T
±0.5 %
> 1.5 T
±1.0 %
5.7.2
AC Field Measurement
In the operating mode AC field measurement the effective value (RMS) is calculated
from the determined AC field contributions. DC field contributions are automatically
suppressed in this type of measurement. The upper display area shows the gathered
DC field contribution. This might be especially helpful when an overflow is displayed
with low AC field amplitude.
Note
Please note that superimposed DC field contributions must not be so
large that they exceed the measuring range. A small AC field
together with a DC field may result in a value
overflow.
The accuracy of the AC field measurement depends on the AC field‘s frequency and
wave shape.
The following shows the accuracies for AC field mesasurements.
Maximum ratings for sinusoidal signal
Frequency
B
eff
Accuracy (1
σ
)
≤ 2 kHz
≤ 1 T
≤ ±1.0 %
≤ 5 kHz
≤ 2 T
≤ ±2.0 %